APICAL FOUR
The Apical Four-Chamber View
TV 6.5 mm nearer the apex
The window behind volumes, ejection fraction, diastolic Doppler and most valve work. This heart is a model, and the details an echo tech checks are in place: the moderator band, the tricuspid valve inserting 6.5 mm nearer the apex than the mitral, and the fossa ovalis dropping out, an artifact of the beam angle, not a defect.
APICAL FOUR
The Method of Disks
20 disks: EF 59 %
Trace the ventricle in two apical views, cut its long axis into twenty slices, each disk taking one diameter from each view. Echo labs call it Simpson’s rule; it is a Riemann sum. Twenty disks give 115.28 mL on a model ventricle of exactly 115.20, and the error falls as the square of the count. A foreshortened apex costs more.
APICAL FOUR
The Strain Bull’s-Eye
GLS −10.7 % · ratio 1.10
Flatten the ventricle onto a disk, apex at the center, and fill each of the 17 segments with peak longitudinal strain. In amyloidosis the base and mid-wall lose strain first: this map averages −10.7%, and its apical average is 1.10 times the basal and mid averages combined, past the ratio of 1 Phelan and colleagues proposed in 2012.
APICAL FOUR
TAPSE
22 mm · S′ 14.5 cm/s
An M-mode cursor through the lateral tricuspid annulus records the base rising toward the apex in systole and falling back in diastole; TAPSE is the height of the hump. On the shirt’s strip it is 22 mm, and its steepest rise gives a peak S′ of 14.5 cm/s. Guidelines call under 17 mm reduced. Quick, reproducible, and one segment of one wall.
APICAL FOUR
Mitral Inflow and the Grades of Diastolic Function
E/A 1.5 · 0.7 · 1.3 · 2.2
Mitral inflow has an early wave, E, as the ventricle relaxes, and a late one, A, when the atrium contracts. The four balances weigh E/A at 1.5, 0.7, 1.3 and 2.2: normal filling, then grades I to III. The third balance is the pseudonormal trap: a healthy-looking ratio at an atrial pressure of 15 mmHg, which tissue Doppler’s e′ exposes.
PARASTERNAL LONG
The Parasternal Long Axis
IVS 0.95 · LVID 4.8 · PW 0.95 cm
The anterior mitral leaflet and the back wall of the aortic root run on as one fibrous sheet, so the view reads as a single structure. The measurements are taken here, at the leaflet tips: in this model 0.95 cm of septum, 4.8 cm of cavity, 0.95 cm of posterior wall. The ring behind the atrium is the descending aorta.
PARASTERNAL LONG
The Continuity Equation
1 mm of LVOT = 10 % of AVA
What goes into a valve comes out of it. Skjaerpe, Hegrenaes and Hatle in 1985 and Zoghbi’s group in 1986 made that the standard measure of aortic stenosis: outflow-tract area times its VTI, divided by the jet’s. Here it gives 0.67 cm², a fifth of the outflow tract. The diameter is squared, so 1 mm in 20 is 10%.
PARASTERNAL LONG
The Vena Contracta
π/(π+2) = 0.611
A jet forced through a sharp-edged hole keeps narrowing after it leaves, because the converging streamlines cannot turn the corner at once. Kirchhoff solved the slot exactly in 1869: the jet settles at π/(π+2), 0.611 of the slot’s width. Hence an effective orifice smaller than the anatomic one, and the empirical coefficient in Gorlin’s formula of 1951.
PARASTERNAL LONG
The M-Mode of the Mitral Valve
EPSS 4.7 mm
One scan line swept through time draws the anterior mitral leaflet as an M: open to the E point in early diastole, drifting nearly shut, reopening at A with atrial contraction, above the posterior leaflet’s smaller W. Here the E point comes 4.7 mm from the septum; a wide gap is a quick sign of a dilated, poorly contracting ventricle.
PARASTERNAL LONG
LV Mass and the Four Geometries
ASE 2015, men: 0.42 · 115 g/m²
Mass says hypertrophy; relative wall thickness says which kind. The four rings are hearts computed from their dimensions and drawn to one scale: normal at 148 g, with 0.9 cm walls around a 4.8 cm cavity; concentric remodeling at 174, thick-walled but not heavy; concentric hypertrophy at 281; eccentric hypertrophy at 283, dilated and thin-walled. The dividing line for relative wall thickness in the 2015 ASE/EACVI guideline is 0.42.
1540
Attenuation and Time-Gain Compensation
0.5 dB/cm/MHz, each way
Soft tissue turns sound into heat at about 0.5 dB per centimeter per megahertz, each way. At 2.5 MHz an echo from 16 cm comes back 40 dB weaker than one from the skin; at 5 MHz, 80 dB. The time-gain sliders undo that by hand, so their diagonal is one straight line with a slope of 2αf.
1540
Phased-Array Steering and the Grating Lobe
pitch λ: a twin beam at −30°
A cardiac probe steers by firing its elements nanoseconds apart, as von Ramm and Thurstone’s first real-time phased-array scanner did at Duke in the early 1970s. The array samples its own aperture: with elements a wavelength apart, a beam steered to +30° grows an equal twin at −30°. Hence the half-wavelength pitch of cardiac arrays.
1540
The Nyquist Limit and Aliasing
E 1.40 m/s reads −0.60
Pulsed Doppler samples the blood once per pulse, and by Nyquist’s theorem of 1928 anything faster than half the sampling rate folds over. At 2.5 MHz and a PRF of 6.5 kHz the limit is 1.00 m/s, so an E wave of 1.40 m/s reads −0.60. In color the core wraps from yellow to cyan with no black between.
1540
Frame Rate and the 77,000 Rule
77,000 ÷ (16 × 96) = 50 Hz
Each line waits 13 µs a centimeter for its echoes, so depth, line count and focal zones share one budget, c/2 = 77,000 cm/s. At 16 cm, 96 lines and one focus give 50 frames a second; the 128-line sector on the monitor above gets 37.6, and a second focal zone halves either. The escapes are on the console: less depth, a narrower sector, fewer lines, or parallel receive beams.
1540
Color Doppler of Mitral Regurgitation
Nyquist 0.60 m/s · jet 8×
Kasai’s autocorrelation estimator of 1985 compares successive echoes from each pixel: red toward the probe, blue away. This jet runs about eight times faster than the 0.60 m/s color scale, so it wraps into a mosaic, while the flow converging on the leak shows its aliasing shell. It covers 46% of the atrium, which is why jet area alone misleads.
LARRD & LATA
Impedance, Gel and the Quarter-Wave Layer
√(33 × 1.63) = 7.3 MRayl
Soft tissue’s acoustic impedance is 1.63 MRayl and air’s about 400 rayl, so a film of air 10 µm thick costs 59 dB: that is the whole case for gel. The crystal, near 33 MRayl, would pass only 18% bare. A quarter-wave layer at the geometric mean, 7.3 MRayl, is the probe’s anti-reflection coat.
LARRD & LATA
Axial Resolution: Half the Pulse
SPL/2 = 0.92 mm at 2.5 MHz
The half comes from the round trip: a reflector Δz deeper sends its echo back 2Δz/c later, and each echo lasts one pulse. A two-cycle pulse at 5 MHz resolves 0.31 mm; computed tone bursts at 2.5 MHz stop overlapping at 0.922 mm, against 0.924 predicted. The backing block keeps pulses short.
LARRD & LATA
Tissue Harmonic Imaging
1.7 MHz out, 3.4 MHz back
Harmonic imaging became routine in echocardiography in the late 1990s and changed the look of the heart. The second harmonic is weak near the skin, where reverberation clutter lives, and grows as the square of the local pressure, so its beam is narrower. Pulse inversion, published in 1999, fires a pulse and its negative: odd harmonics cancel, the second doubles.
LARRD & LATA
The Beam: Near Zone, Focus, Far Zone
N = D²/4λ = 61.6 mm · 3.5°
Near an unfocused disk, wavelets from across the face interfere and the axis swings between maxima and zeros: the Fresnel zone. The last maximum, the natural focus, lies near D²/4λ, 61.6 mm for a 12.3 mm disk at 2.5 MHz, where the beam is 5.1 mm wide at −6 dB. Beyond it the beam spreads at 3.5°.
LARRD & LATA
Speckle
mean/sd = 1.91, any tissue
Each resolution cell holds many scatterers far smaller than a wavelength, returning echoes of random phase, and their sum is a random walk whose length follows Rayleigh’s distribution. The simulation behind this phantom gives a mean-to-standard-deviation ratio of 1.908 against the theoretical 1.91. The grain is noise, but stable noise that moves with the muscle: speckle tracking depends on it.
FOUR V SQUARED
PISA: the Proximal Isovelocity Surface Area
r 0.83 cm at 38.5 cm/s
Blood accelerating toward a leak passes through shells of equal speed, and color Doppler draws one where it aliases. Its radius gives the flow, 2πr² times the aliasing velocity, and flow over jet velocity gives the orifice. Here the radius is 0.83 cm, and the hemisphere formula reads 0.33 cm² for a true 0.40: near the orifice the shells flatten.
FOUR V SQUARED
Pressure Half-Time and the Number 220
220 ms ↔ 1.0 cm²
Hatle’s rule: mitral area is 220 over the half-time, so 220 ms means 1 cm², the area of this orifice. Pressure goes as velocity squared, so it halves when velocity falls by √2. In a typical heart 220 is 11.6 times net compliance times the root of the initial gradient: a compliance in disguise, so it fails after balloon valvotomy.
FOUR V SQUARED
RVSP = 4v² + RAP
ASE 2010: 3 · 8 · 15 mmHg
Yock and Popp showed in 1984 that the peak tricuspid regurgitant velocity gives the pressure drop from right ventricle to right atrium. Add atrial pressure, read from the inferior vena cava: at most 2.1 cm and collapsing by more than half with a sniff, 3 mmHg; wider and collapsing less, 15; anything else, 8.
FOUR V SQUARED
The Dome and the Dagger
peak at 39 % vs 60 %
From the apex, continuous-wave Doppler records a fast systolic jet in two diseases, and the shape tells them apart. Fixed aortic stenosis makes a rounded dome. Dynamic obstruction in hypertrophic cardiomyopathy makes a dagger: the outflow tract narrows while blood is still leaving, so velocity peaks late, at 60% of ejection here against 39% for the dome.
FOUR V SQUARED
dP/dt from the Mitral Jet
4(3² − 1²) = 32 mmHg
While a mitral regurgitant jet speeds from 1 to 3 m/s, atrial pressure barely moves, so the ventricle has added 4(9 − 1) = 32 mmHg; divide by the time taken. Bargiggia and colleagues described it in 1989. In the model a brisk ventricle crosses the window in about 22 ms, near 1,400 mmHg/s; a depressed one takes 64 ms, about 500.