Cardiac Sonography · Subject 09

Cardiac Sonography

Button-ups, ties and socks drawn from the views, jets and equations you scan by.

Apical four-chamber · 16 cm · 128 lines · 37.6 frames/s

This week in Cardiac Sonography,

A guideline Doppler cutoff for tricuspid leaks rarely trips, an AI reads Takotsubo in diastole, low forward flow hides behind normal ejection fraction, and ultrasound maps microvessels in beating rat hearts.

Get it by email

Download MP3

Behind this page an echo machine runs a full transthoracic study of one model patient, the same heart as the Cardiology page: parasternal long and short axis, the apical four-, two- and three-chamber views and the subcostal views, with M-mode, color, pulsed and tissue Doppler and continuous wave. Each view is measured as a sonographer would: the ejection fraction by the biplane method of disks is 58 %, mitral inflow E is 0.83 m/s, e′ is 10 cm/s at the septum, and the report fills in as the page goes on.

Welcome in

For anyone who has traced an envelope

You know the ache in your scanning arm by the last study of the day, the gel warmer, the dimmed room, and the move one rib space down that finally stops the apex foreshortening. You know the heart upside down, apex at the top of the fan and left ventricle on the right, better than any textbook heart the right way up. You remember the first Doppler envelope you traced that gave a number you believed: a jet at 4 m/s, a gradient of 64 mmHg.

Start with everything

ALL VIEWS, Bone
ALL VIEWSRegular price $65 Sale price $58.50Shop now

Colorway
Bone

Whole-field medallions · repeats every 18 in

Short-sleeve button-up · XS–4XL

ALL VIEWS

The whole field, on one shirt.

Cardiac Sonography brought together in one design, printed edge to edge: 25 medallions from the field’s designs, with the equations from across the subject. Related images are spread across the pattern so no two neighbors repeat.

  • Print scaleThe pattern repeats every 18 inches, with seamless edges.
  • StylesPhotoreal on bone, engraving or blueprint.
Shop now · Regular price $65 Sale price $58.50

Leave your email and we’ll write the day it’s ready.

ECHO TERRAIN, Gel
ECHO TERRAINRegular price $65 Sale price $58.50Shop now

Colorway

Contour map · repeats every 18 in

Short-sleeve button-up · XS–4XL

ECHO TERRAIN

A topographic portrait of the field.

Recognizable forms from Cardiac Sonography rise as contour landscapes. Lines follow the outlines and continue across the ground between them, with names fitted along the contours.

  • Print scaleThe pattern repeats every 18 inches, with seamless edges.
  • The paletteTwo colorways, shown at the same print scale.
Shop now · Regular price $65 Sale price $58.50

Leave your email and we’ll write the day it’s ready.

Socks mix and match across every field: three pairs for $69, and two ship free in the U.S.

Crew socks · One size

Crew socks

A field of your own, from cuff to toe.

  • The pattern
  • The previewA studio mockup; photos of the finished pair are on the way.
Shop now · Regular price $26 Sale price $23.40

Leave your email and we’ll write the day this pair is ready.

Necktie · 3.5 in wide

Subject ties

A formal introduction to your field.

  • The color
  • The previewStudio render of the tie. Zoom shows the print up close.
Shop now · Regular price $36 Sale price $32.40

Leave your email and we’ll write the day this tie is ready.

More designs to explore

Short-sleeve button-up · XS–4XL

PARASTERNAL LONG

The first image of every study, and the numbers that start there.

Every transthoracic study starts beside the left edge of the sternum, in the third or fourth intercostal space, with the beam slicing the heart along its long axis. The right ventricle lies across the top, then the septum, the left ventricular cavity and the posterior wall, with the aortic root and left atrium on the right. The anterior mitral leaflet runs straight on from the back wall of the aorta, and the small circle behind the left atrium is the descending aorta: pericardial fluid tracks in front of it, pleural fluid behind. The M-mode through the mitral leaflet tips predates two-dimensional echo: the first echocardiogram was an M-mode, recorded in Lund in 1953 by the cardiologist Inge Edler and the physicist Hellmuth Hertz with an industrial flaw detector built to find cracks in metal.

Much of the report starts in this view. Septum, cavity and posterior wall at end-diastole give left ventricular mass by the cube formula, the shell between two cubes times the density of muscle, 1.04 g/mL, corrected against necropsy by Devereux in 1986, and in 1992 Ganau and colleagues crossed mass with relative wall thickness to sort hearts into four geometries. The outflow-tract diameter measured here feeds the continuity equation for aortic stenosis, where it is squared, so 1 mm of error in 20 becomes 10% in the valve area. And the neck of a regurgitant jet, the vena contracta, is narrower than the hole it came through: for a slot, Kirchhoff’s exact solution of 1869 gives π/(π+2), 0.611 of the width.

  • The Parasternal Long AxisThe long-axis view at end-diastole: septum, cavity and posterior wall, aortic root and left atrium, and the small ring of descending aorta behind.
  • The Continuity EquationA calcified tricuspid aortic valve in the bronze of 3-D echo, open to a small three-cornered orifice: a fifth of the outflow tract.
  • The Vena ContractaSilver wires converging on a slot in a steel plate, two copper free streamlines narrowing the jet below to 0.611 of the slot.
  • The M-Mode of the Mitral ValveTwo beats through the leaflet tips: the anterior leaflet’s M, the posterior leaflet’s smaller W, and the E point 4.7 mm from the septum.
  • LV Mass and the Four GeometriesFour oxblood porcelain rings to one scale around a gold cross: normal, concentric remodeling, concentric and eccentric hypertrophy, as Ganau sorted them.
Shop now · Regular price $65 Sale price $58.50

Leave your email and we’ll write the day it’s ready.

Short-sleeve button-up · XS–4XL

APICAL FOUR

Probe on the apex, heart upside down: the window most function numbers come from.

Put the probe on the apex beat, aim at the right shoulder, and the heart appears upside down: apex at the top, because the top of the image is always nearest the transducer, and, by the display convention standardized in 1980, the left ventricle on the right of the screen. All four chambers, both atrioventricular valves and both septa show at once, so most of the function page of a report is measured here. The guidelines’ method for volumes and ejection fraction is the method of disks, which echo labs call Simpson’s rule. It is not: Thomas Simpson’s rule of 1743 fits parabolas, while the echo method traces two apical views and stacks twenty elliptical disks, a plain Riemann sum.

The right ventricle defeats that geometry. It is a crescent wrapped around the left and empties mostly by pulling its base toward the apex, so in 1984 Kaul and colleagues proposed measuring how far the base moves: TAPSE, one cursor and one caliper. Speckle tracking made strain routine. The American Heart Association’s 17 segments of 2002 flatten into a bull’s-eye of peak longitudinal strain, about −20% in health, and when the base and mid-wall fade while the apex keeps its color, sonographers say “cherry on top” before they say amyloid. Diastole has its own trap. Appleton, Hatle and Popp matched the inflow patterns to catheter pressures in 1988, and a pseudonormal E/A looks healthy at a far higher atrial pressure; tissue Doppler’s e′ breaks the tie.

  • The Apical Four-Chamber ViewA grayscale fan, point up: the heart upside down, left ventricle on the right, the moderator band crossing the right ventricular apex.
  • The Method of DisksTwenty glass disks stacked apex-up into a left ventricle, widest two thirds of the way down: the Riemann sum echo labs call Simpson’s rule.
  • The Strain Bull’s-EyeCloisonné enamel in 17 cells, pale shell-pink at the rim deepening to cherry at the center: the apical sparing of amyloid.
  • TAPSEAn M-mode strip two beats long: the tricuspid annulus rising steeply 22 mm in systole and settling back more slowly through diastole.
  • Mitral Inflow and the Grades of Diastolic FunctionFour brass balances weighing E against A: down left, down right, down left again, then steeply left. The third one is the trap.
Shop now · Regular price $65 Sale price $58.50

Leave your email and we’ll write the day it’s ready.

Short-sleeve button-up · XS–4XL

1540

1540 m/s, the speed of sound in soft tissue, and every limit it sets on the console.

Every ultrasound machine assumes that sound crosses soft tissue at 1540 meters a second, and a surprising amount of the console follows from that one number. An echo from 1 cm deep is home in 13 µs and the next pulse must wait for it, so a 16 cm sector of 96 lines takes 19.9 ms, which caps it at fifty frames a second. Deeper echoes also come back weaker. Tissue absorbs about 0.5 dB per centimeter per megahertz each way, so at 2.5 MHz an echo from 16 cm returns 40 dB down, and the bank of time-gain sliders is that loss undone by hand.

Speed over frequency is wavelength, 0.616 mm at 2.5 MHz, and it sets the rest. A phased array steers by firing its elements nanoseconds apart, as in the first real-time phased-array cardiac scanner, built by von Ramm and Thurstone at Duke in the early 1970s; space the elements a whole wavelength apart and a phantom beam of equal strength appears on the other side, which is why cardiac arrays are built at half a wavelength. The same wait caps the pulse rate and makes fast blood alias, and depth and speed trade against each other: at 10 cm, pulsed Doppler at 2.5 MHz reads no faster than 1.19 m/s, which is why deep, fast jets need continuous wave.

  • Attenuation and Time-Gain CompensationA brushed-aluminum bank of eight sliders, the knobs on one straight diagonal: 40 dB of extra gain by 16 cm at 2.5 MHz.
  • Phased-Array Steering and the Grating LobeA phased-array probe steered 30° right with its elements a wavelength apart, and the phantom twin beam, just as bright, 30° left.
  • The Nyquist Limit and AliasingZoomed color Doppler of mitral inflow: red brightening to yellow, then a cyan core with no black between. Aliasing, not reversal.
  • Frame Rate and the 77,000 RuleNinety-six steel wires across a quarter circle, one per scan line, enameled blue to amber in firing order: 19.9 ms a frame.
  • Color Doppler of Mitral RegurgitationThe four-chamber view in systole: a blue convergence dome with a yellow aliasing rim, and a mosaic jet filling half the left atrium.
Shop now · Regular price $65 Sale price $58.50

Leave your email and we’ll write the day it’s ready.

Short-sleeve button-up · XS–4XL

LARRD & LATA

Resolution the way the registry teaches it: along the beam and across it.

Everyone who sat the ultrasound physics registry chanted two mnemonics: LARRD for resolution along the beam, longitudinal, axial, range, radial and depth, and LATA for resolution across it. Along the beam, two reflectors show as two only if their echoes do not overlap, and the deeper echo lags by its extra round trip, so axial resolution is half the spatial pulse length: at 2.5 MHz a three-cycle pulse is 1.85 mm long and resolves 0.92 mm. Short pulses need a heavily damped crystal and a wide bandwidth. The crystal sits near 33 MRayl against tissue’s 1.63; one quarter-wave matching layer bridges the gap, and a second widens the half-power band from 62% to 96% of the center frequency.

Across the beam, resolution is the width of the beam. An unfocused disk interferes with itself in the near zone, narrows to its natural focus and then spreads at an angle set by 1.22 λ/D, the number that limits a telescope; a wider aperture or a higher frequency pushes the focus deeper and narrows the far beam. Tissue harmonic imaging narrows the beam further. Sound runs slightly faster through compressed tissue than rarefied, so a strong wave steepens and the tissue itself makes a second harmonic, 3.4 MHz from 1.7, with a narrower beam and lower side lobes. And the grain of the image is the resolution cell itself: Burckhardt showed in 1978 that fully developed speckle has a mean-to-standard-deviation ratio of 1.91, whatever the tissue.

  • Impedance, Gel and the Quarter-Wave LayerA cutaway probe: bubble-free gel, lens, two quarter-wave matching layers, a diced crystal and the dark backing block that stops it ringing.
  • Axial Resolution: Half the PulseSchlieren pulses in three channels: echoes that merge, echoes from wires half a pulse length apart that just touch, and echoes with a gap.
  • Tissue Harmonic ImagingOne short-axis doughnut on a split screen: fundamental on the left, hazy and coarse; second harmonic on the right, clean and crisp.
  • The Beam: Near Zone, Focus, Far ZoneA disk transducer’s beam in golden schlieren light: fringes and beads out to the natural focus at 61.6 mm, then a slim 3.5° spread.
  • SpeckleA quality-control phantom: uniform speckle, a column of six wire targets each the size of one grain, and three echo-free cysts.
Shop now · Regular price $65 Sale price $58.50

Leave your email and we’ll write the day it’s ready.

Short-sleeve button-up · XS–4XL

FOUR V SQUARED

Velocities in, pressures and orifices out: the Doppler numbers on every report.

Bernoulli’s equation, cut down to four times the velocity squared, turns a Doppler velocity into a pressure difference, and many of the hemodynamic numbers on an echo report come out of it. Yock and Popp showed in 1984 that the peak velocity of tricuspid regurgitation gives the pressure drop from right ventricle to right atrium; add an atrial pressure read from the inferior vena cava by its width and its collapse with a sniff, 3, 8 or 15 mmHg in the ASE scheme of 2010, and the sum is right ventricular systolic pressure. The same arithmetic times a ventricle. While a mitral jet speeds from 1 to 3 m/s the ventricle adds 4(9 − 1) = 32 mmHg, so the time taken gives dP/dt, as Bargiggia and colleagues showed in 1989.

Flow is conserved too. Blood converging on a leaking valve passes through shells of equal speed, and color Doppler draws one of them where it aliases; Bargiggia and colleagues brought the hemisphere formula to the clinic in 1991, and it reads low near the orifice, where the shells flatten. In 1979 Liv Hatle and colleagues noticed that the tighter a stenotic mitral valve, the longer its gradient takes to halve, and the constant 220 was chosen so that 220 ms means 1 cm². And continuous-wave Doppler from the apex tells two fast systolic jets apart by their shape: the rounded dome of fixed aortic stenosis, and the late-peaking dagger of dynamic obstruction in hypertrophic cardiomyopathy.

  • PISA: the Proximal Isovelocity Surface AreaNested glass shells over a leak: blue hemispheres outside, one thin yellow aliasing shell, and red shells flattening against the orifice.
  • Pressure Half-Time and the Number 220A rheumatic mitral valve in the bronze of 3-D echo, seen from the atrium: fused commissures and a fish-mouth orifice of 1 cm².
  • RVSP = 4v² + RAPA subcostal view of the inferior vena cava entering the right atrium, calipers across it: the pressure that gets added to 4v².
  • The Dome and the DaggerTwo silver pendants cast from continuous-wave envelopes: the rounded dome of aortic stenosis and the late, hollow-edged dagger of obstruction.
  • dP/dt from the Mitral JetTwo sand-glasses started together, each holding 32 mmHg of sand: the brisk ventricle has emptied, the depressed one is a third through.
Shop now · Regular price $65 Sale price $58.50

Leave your email and we’ll write the day it’s ready.

Short-sleeve button-up · XS–4XL

EQUATION STRIPE

From acoustic impedance to the continuity equation, every line labeled.

Cardiac sonography runs on a few constants and a great deal of squaring: 1540 m/s, 13 µs a centimeter, 0.5 dB per centimeter per megahertz, four times the velocity squared. The Equation Stripe sets the equations behind the collection in rows, each introduced by its name in capitals, so SIMPLIFIED BERNOULLI, CONTINUITY EQUATION and PRESSURE HALF-TIME read at a glance. Together they follow a study from the probe face to the report: how the pulse is made, focused and steered, how the image forms and why it is grainy, how velocity becomes pressure and orifice area, and how volume, mass and strain are measured.

Every row carries the science of the collection’s five picture designs. Some are physics learned for the registry, others measurements made every working day. A sonographer will know most of them by their constants before reading a symbol: 220 for the mitral valve area, 1.04 g/mL in the mass formula and twenty disks in the volume.

  • Acoustic ImpedanceZ = ρc = 1.63 MRayl (soft tissue)
  • Intensity ReflectionR = ((Z2 − Z1)/(Z2 + Z1))2 ≈ 99.9% (tissue → air)
  • Quarter-Wave MatchingZm = √(ZpZt) ≈ 7.3 MRayl, d = λm/4
  • Natural FocusN = D2/(4λ) = 61.6 mm (D = 20λ, 2.5 MHz)
  • Beam Divergencesin θ = 1.22λ/D: D = 20λ → θ = 3.50°
  • Steering Delaysτn = np sin θs/c
Shop now · Regular price $65 Sale price $58.50

Leave your email and we’ll write the day it’s ready.

Short-sleeve button-up · XS–4XL

CONCEPT STRIPE

From the parasternal long axis to the bubble study, the whole scan in short lines.

The Concept Stripe carries cardiac sonography’s working knowledge: the views and how to find them, the measurements and what they mean, the Doppler physics and the signs, each after a short heading in capitals; guideline thresholds appear as reference values, attributed where the societies differ. About every fourth unit is an equation from the Equation Stripe, so SIMPLIFIED BERNOULLI and the CONTINUITY EQUATION turn up between PLAX and the McConnell sign.

  • PLAX3rd–4th space by the sternum, marker to right shoulder
  • LVEFnormal 52–72% in men, 54–74% in women (ASE/EACVI)
  • INSONATION ANGLEunder 20°: cos 20° = 0.94, at most a 6% error
  • PISA SET-UPzoom in, shift the color baseline to 30–40 cm/s
  • PSAXturn 90° clockwise from PLAX, marker to left shoulder
  • LA VOLUME INDEXnormal ≤ 34 mL/m², biplane at LV end-systole
Shop now · Regular price $65 Sale price $58.50

Leave your email and we’ll write the day it’s ready.

ACOUSTIC IMPEDANCEZ = ρc = 1.63 MRayl (soft tissue)·INTENSITY REFLECTIONR = ((Z2 − Z1)/(Z2 + Z1))2 ≈ 99.9% (tissue → air)·QUARTER-WAVE MATCHINGZm = √(ZpZt) ≈ 7.3 MRayl, d = λm/4·NATURAL FOCUSN = D2/(4λ) = 61.6 mm (D = 20λ, 2.5 MHz)·BEAM DIVERGENCEsin θ = 1.22λ/D: D = 20λ → θ = 3.50°·STEERING DELAYSτn = np sin θs/c·GRATING LOBEsin θg = sin θs − mλ/p: p = λ, θs = 30° → θg = −30°·ELEMENT PITCHp < λ/(1 + |sin θs|) = 0.586λ at ±45°·ATTENUATIONL = 2αfd, α ≈ 0.5 dB cm−1MHz−1·TIME-GAIN COMPENSATIONG(t) = 2αf(ct/2) = αfct·AXIAL RESOLUTIONΔz = SPL/2 = nλ/2 = nc/(2f)·FRAME RATEFR = (c/2)/(dNlinesNfoci), c/2 = 77,000 cm s−1·NYQUIST LIMITvNyq = c·PRF/(4f0cos θ)·RANGE–VELOCITY LIMITvmax = c2/(8f0d): 1.19 m s−1 at 10 cm, 2.5 MHz·RANGE EQUATIONr = ct/2, c = 1540 m s−1 ⇒ 2/c = 13 μs cm−1·B-MODE FORMATIONI(r,θ) = 20log10|h ∗ (aξ)|·RAYLEIGH SCATTERINGσs ∝ k4a6 ∝ f4 (ka ≪ 1)·HARMONIC GENERATION∂p/∂z = (β/(ρc3))p ∂p/∂τ, β = 1 + B/(2A)·FUBINI SOLUTIONpn = (2p0/(nσ))Jn(nσ), σ = z/x̄, x̄ = ρc3/(βωp0)·SPECKLE STATISTICSp(A) = (A/σ2)exp(−A2/(2σ2))·SPECKLE SNR⟨A⟩/σA = √(π/(4 − π)) = 1.91·COLOR AUTOCORRELATIONv̄ = (c·PRF/(4πf0)) arg R(1)·LAG-ONE CORRELATIONR(1) = (1/(N − 1))∑n=1N−1sn+1sn*·DOPPLER EQUATIONv = cfD/(2f0cos θ)·ACOUSTIC IMPEDANCEZ = ρc = 1.63 MRayl (soft tissue)·INTENSITY REFLECTIONR = ((Z2 − Z1)/(Z2 + Z1))2 ≈ 99.9% (tissue → air)·QUARTER-WAVE MATCHINGZm = √(ZpZt) ≈ 7.3 MRayl, d = λm/4·NATURAL FOCUSN = D2/(4λ) = 61.6 mm (D = 20λ, 2.5 MHz)·BEAM DIVERGENCEsin θ = 1.22λ/D: D = 20λ → θ = 3.50°·STEERING DELAYSτn = np sin θs/c·GRATING LOBEsin θg = sin θs − mλ/p: p = λ, θs = 30° → θg = −30°·ELEMENT PITCHp < λ/(1 + |sin θs|) = 0.586λ at ±45°·ATTENUATIONL = 2αfd, α ≈ 0.5 dB cm−1MHz−1·TIME-GAIN COMPENSATIONG(t) = 2αf(ct/2) = αfct·AXIAL RESOLUTIONΔz = SPL/2 = nλ/2 = nc/(2f)·FRAME RATEFR = (c/2)/(dNlinesNfoci), c/2 = 77,000 cm s−1·NYQUIST LIMITvNyq = c·PRF/(4f0cos θ)·RANGE–VELOCITY LIMITvmax = c2/(8f0d): 1.19 m s−1 at 10 cm, 2.5 MHz·RANGE EQUATIONr = ct/2, c = 1540 m s−1 ⇒ 2/c = 13 μs cm−1·B-MODE FORMATIONI(r,θ) = 20log10|h ∗ (aξ)|·RAYLEIGH SCATTERINGσs ∝ k4a6 ∝ f4 (ka ≪ 1)·HARMONIC GENERATION∂p/∂z = (β/(ρc3))p ∂p/∂τ, β = 1 + B/(2A)·FUBINI SOLUTIONpn = (2p0/(nσ))Jn(nσ), σ = z/x̄, x̄ = ρc3/(βωp0)·SPECKLE STATISTICSp(A) = (A/σ2)exp(−A2/(2σ2))·SPECKLE SNR⟨A⟩/σA = √(π/(4 − π)) = 1.91·COLOR AUTOCORRELATIONv̄ = (c·PRF/(4πf0)) arg R(1)·LAG-ONE CORRELATIONR(1) = (1/(N − 1))∑n=1N−1sn+1sn*·DOPPLER EQUATIONv = cfD/(2f0cos θ)·

The science on the shirt

What you’re wearing

Every object on the shirts, and the science behind it.

The Apical Four-Chamber View

APICAL FOUR

The Apical Four-Chamber View

TV 6.5 mm nearer the apex

The window behind volumes, ejection fraction, diastolic Doppler and most valve work. This heart is a model, and the details an echo tech checks are in place: the moderator band, the tricuspid valve inserting 6.5 mm nearer the apex than the mitral, and the fossa ovalis dropping out, an artifact of the beam angle, not a defect.

The Method of Disks

APICAL FOUR

The Method of Disks

20 disks: EF 59 %

Trace the ventricle in two apical views, cut its long axis into twenty slices, each disk taking one diameter from each view. Echo labs call it Simpson’s rule; it is a Riemann sum. Twenty disks give 115.28 mL on a model ventricle of exactly 115.20, and the error falls as the square of the count. A foreshortened apex costs more.

The Strain Bull’s-Eye

APICAL FOUR

The Strain Bull’s-Eye

GLS −10.7 % · ratio 1.10

Flatten the ventricle onto a disk, apex at the center, and fill each of the 17 segments with peak longitudinal strain. In amyloidosis the base and mid-wall lose strain first: this map averages −10.7%, and its apical average is 1.10 times the basal and mid averages combined, past the ratio of 1 Phelan and colleagues proposed in 2012.

TAPSE

APICAL FOUR

TAPSE

22 mm · S′ 14.5 cm/s

An M-mode cursor through the lateral tricuspid annulus records the base rising toward the apex in systole and falling back in diastole; TAPSE is the height of the hump. On the shirt’s strip it is 22 mm, and its steepest rise gives a peak S′ of 14.5 cm/s. Guidelines call under 17 mm reduced. Quick, reproducible, and one segment of one wall.

Mitral Inflow and the Grades of Diastolic Function

APICAL FOUR

Mitral Inflow and the Grades of Diastolic Function

E/A 1.5 · 0.7 · 1.3 · 2.2

Mitral inflow has an early wave, E, as the ventricle relaxes, and a late one, A, when the atrium contracts. The four balances weigh E/A at 1.5, 0.7, 1.3 and 2.2: normal filling, then grades I to III. The third balance is the pseudonormal trap: a healthy-looking ratio at an atrial pressure of 15 mmHg, which tissue Doppler’s e′ exposes.

The Parasternal Long Axis

PARASTERNAL LONG

The Parasternal Long Axis

IVS 0.95 · LVID 4.8 · PW 0.95 cm

The anterior mitral leaflet and the back wall of the aortic root run on as one fibrous sheet, so the view reads as a single structure. The measurements are taken here, at the leaflet tips: in this model 0.95 cm of septum, 4.8 cm of cavity, 0.95 cm of posterior wall. The ring behind the atrium is the descending aorta.

The Continuity Equation

PARASTERNAL LONG

The Continuity Equation

1 mm of LVOT = 10 % of AVA

What goes into a valve comes out of it. Skjaerpe, Hegrenaes and Hatle in 1985 and Zoghbi’s group in 1986 made that the standard measure of aortic stenosis: outflow-tract area times its VTI, divided by the jet’s. Here it gives 0.67 cm², a fifth of the outflow tract. The diameter is squared, so 1 mm in 20 is 10%.

The Vena Contracta

PARASTERNAL LONG

The Vena Contracta

π/(π+2) = 0.611

A jet forced through a sharp-edged hole keeps narrowing after it leaves, because the converging streamlines cannot turn the corner at once. Kirchhoff solved the slot exactly in 1869: the jet settles at π/(π+2), 0.611 of the slot’s width. Hence an effective orifice smaller than the anatomic one, and the empirical coefficient in Gorlin’s formula of 1951.

The M-Mode of the Mitral Valve

PARASTERNAL LONG

The M-Mode of the Mitral Valve

EPSS 4.7 mm

One scan line swept through time draws the anterior mitral leaflet as an M: open to the E point in early diastole, drifting nearly shut, reopening at A with atrial contraction, above the posterior leaflet’s smaller W. Here the E point comes 4.7 mm from the septum; a wide gap is a quick sign of a dilated, poorly contracting ventricle.

LV Mass and the Four Geometries

PARASTERNAL LONG

LV Mass and the Four Geometries

ASE 2015, men: 0.42 · 115 g/m²

Mass says hypertrophy; relative wall thickness says which kind. The four rings are hearts computed from their dimensions and drawn to one scale: normal at 148 g, with 0.9 cm walls around a 4.8 cm cavity; concentric remodeling at 174, thick-walled but not heavy; concentric hypertrophy at 281; eccentric hypertrophy at 283, dilated and thin-walled. The dividing line for relative wall thickness in the 2015 ASE/EACVI guideline is 0.42.

Attenuation and Time-Gain Compensation

1540

Attenuation and Time-Gain Compensation

0.5 dB/cm/MHz, each way

Soft tissue turns sound into heat at about 0.5 dB per centimeter per megahertz, each way. At 2.5 MHz an echo from 16 cm comes back 40 dB weaker than one from the skin; at 5 MHz, 80 dB. The time-gain sliders undo that by hand, so their diagonal is one straight line with a slope of 2αf.

Phased-Array Steering and the Grating Lobe

1540

Phased-Array Steering and the Grating Lobe

pitch λ: a twin beam at −30°

A cardiac probe steers by firing its elements nanoseconds apart, as von Ramm and Thurstone’s first real-time phased-array scanner did at Duke in the early 1970s. The array samples its own aperture: with elements a wavelength apart, a beam steered to +30° grows an equal twin at −30°. Hence the half-wavelength pitch of cardiac arrays.

The Nyquist Limit and Aliasing

1540

The Nyquist Limit and Aliasing

E 1.40 m/s reads −0.60

Pulsed Doppler samples the blood once per pulse, and by Nyquist’s theorem of 1928 anything faster than half the sampling rate folds over. At 2.5 MHz and a PRF of 6.5 kHz the limit is 1.00 m/s, so an E wave of 1.40 m/s reads −0.60. In color the core wraps from yellow to cyan with no black between.

Frame Rate and the 77,000 Rule

1540

Frame Rate and the 77,000 Rule

77,000 ÷ (16 × 96) = 50 Hz

Each line waits 13 µs a centimeter for its echoes, so depth, line count and focal zones share one budget, c/2 = 77,000 cm/s. At 16 cm, 96 lines and one focus give 50 frames a second; the 128-line sector on the monitor above gets 37.6, and a second focal zone halves either. The escapes are on the console: less depth, a narrower sector, fewer lines, or parallel receive beams.

Color Doppler of Mitral Regurgitation

1540

Color Doppler of Mitral Regurgitation

Nyquist 0.60 m/s · jet 8×

Kasai’s autocorrelation estimator of 1985 compares successive echoes from each pixel: red toward the probe, blue away. This jet runs about eight times faster than the 0.60 m/s color scale, so it wraps into a mosaic, while the flow converging on the leak shows its aliasing shell. It covers 46% of the atrium, which is why jet area alone misleads.

Impedance, Gel and the Quarter-Wave Layer

LARRD & LATA

Impedance, Gel and the Quarter-Wave Layer

√(33 × 1.63) = 7.3 MRayl

Soft tissue’s acoustic impedance is 1.63 MRayl and air’s about 400 rayl, so a film of air 10 µm thick costs 59 dB: that is the whole case for gel. The crystal, near 33 MRayl, would pass only 18% bare. A quarter-wave layer at the geometric mean, 7.3 MRayl, is the probe’s anti-reflection coat.

Axial Resolution: Half the Pulse

LARRD & LATA

Axial Resolution: Half the Pulse

SPL/2 = 0.92 mm at 2.5 MHz

The half comes from the round trip: a reflector Δz deeper sends its echo back 2Δz/c later, and each echo lasts one pulse. A two-cycle pulse at 5 MHz resolves 0.31 mm; computed tone bursts at 2.5 MHz stop overlapping at 0.922 mm, against 0.924 predicted. The backing block keeps pulses short.

Tissue Harmonic Imaging

LARRD & LATA

Tissue Harmonic Imaging

1.7 MHz out, 3.4 MHz back

Harmonic imaging became routine in echocardiography in the late 1990s and changed the look of the heart. The second harmonic is weak near the skin, where reverberation clutter lives, and grows as the square of the local pressure, so its beam is narrower. Pulse inversion, published in 1999, fires a pulse and its negative: odd harmonics cancel, the second doubles.

The Beam: Near Zone, Focus, Far Zone

LARRD & LATA

The Beam: Near Zone, Focus, Far Zone

N = D²/4λ = 61.6 mm · 3.5°

Near an unfocused disk, wavelets from across the face interfere and the axis swings between maxima and zeros: the Fresnel zone. The last maximum, the natural focus, lies near D²/4λ, 61.6 mm for a 12.3 mm disk at 2.5 MHz, where the beam is 5.1 mm wide at −6 dB. Beyond it the beam spreads at 3.5°.

Speckle

LARRD & LATA

Speckle

mean/sd = 1.91, any tissue

Each resolution cell holds many scatterers far smaller than a wavelength, returning echoes of random phase, and their sum is a random walk whose length follows Rayleigh’s distribution. The simulation behind this phantom gives a mean-to-standard-deviation ratio of 1.908 against the theoretical 1.91. The grain is noise, but stable noise that moves with the muscle: speckle tracking depends on it.

PISA: the Proximal Isovelocity Surface Area

FOUR V SQUARED

PISA: the Proximal Isovelocity Surface Area

r 0.83 cm at 38.5 cm/s

Blood accelerating toward a leak passes through shells of equal speed, and color Doppler draws one where it aliases. Its radius gives the flow, 2πr² times the aliasing velocity, and flow over jet velocity gives the orifice. Here the radius is 0.83 cm, and the hemisphere formula reads 0.33 cm² for a true 0.40: near the orifice the shells flatten.

Pressure Half-Time and the Number 220

FOUR V SQUARED

Pressure Half-Time and the Number 220

220 ms ↔ 1.0 cm²

Hatle’s rule: mitral area is 220 over the half-time, so 220 ms means 1 cm², the area of this orifice. Pressure goes as velocity squared, so it halves when velocity falls by √2. In a typical heart 220 is 11.6 times net compliance times the root of the initial gradient: a compliance in disguise, so it fails after balloon valvotomy.

RVSP = 4v² + RAP

FOUR V SQUARED

RVSP = 4v² + RAP

ASE 2010: 3 · 8 · 15 mmHg

Yock and Popp showed in 1984 that the peak tricuspid regurgitant velocity gives the pressure drop from right ventricle to right atrium. Add atrial pressure, read from the inferior vena cava: at most 2.1 cm and collapsing by more than half with a sniff, 3 mmHg; wider and collapsing less, 15; anything else, 8.

The Dome and the Dagger

FOUR V SQUARED

The Dome and the Dagger

peak at 39 % vs 60 %

From the apex, continuous-wave Doppler records a fast systolic jet in two diseases, and the shape tells them apart. Fixed aortic stenosis makes a rounded dome. Dynamic obstruction in hypertrophic cardiomyopathy makes a dagger: the outflow tract narrows while blood is still leaving, so velocity peaks late, at 60% of ejection here against 39% for the dome.

dP/dt from the Mitral Jet

FOUR V SQUARED

dP/dt from the Mitral Jet

4(3² − 1²) = 32 mmHg

While a mitral regurgitant jet speeds from 1 to 3 m/s, atrial pressure barely moves, so the ventricle has added 4(9 − 1) = 32 mmHg; divide by the time taken. Bargiggia and colleagues described it in 1989. In the model a brisk ventricle crosses the window in about 22 ms, near 1,400 mmHg/s; a depressed one takes 64 ms, about 500.